Abstract Background: The Aspergillus Galactomannan Ag Virclia® (GMVClia) test is a monotest and automated galactomannan technique based on chemiluminescent immunoassay (CLIA). Aim: To evaluate the performance of the GM-VClia test in serum and bronchioalveolar lavage (BAL) samples previously processed with the Platelia ™ Aspergillus EIA kit (GM-Plat). Methods: 56 samples of serum 40 from BAL (some of them with galactomaman determination in both samples), from patients with pulmonary diseases, hematological diseases, SLE, Covid-19 and tumors, among others, were studied. Thirteen patients had invasive aspergillosis (1 proven and 12 probable). Results: The correlation between both methods for serum and BAL was r = 0.8861 p < 0.0001 and r = 0.6368 p < 0.001, respectively. There was a global concordance of 67.7% (65/96), being 85.7% (48/56) in sera and 42.5.0% (14/49) in BAL. By raising the cut-off point in LBA by GM-VClia, the agreement increased to 85.7%. Conclusion: A greater correlation and concordance was observed in sera than in BAL. The GM-VClia kit had a higher sensitivity and NPV than the GM-Plat kit. The disadvantages of GM-VClia are the “doubtful” category, which makes interpretation difficult and that with the current cut-off points in LBA the correlation with GM-Plat is lower. The advantages are its greater sensitivity, ease of processing and faster results.
Resumen Introducción: La prueba Aspergillus galactomannan Ag Virclia® (GM-VClia) es una técnica de galactomanano monotest, auto-matizada, basada en inmunoensayo quimioluminiscente (CLIA). Objetivo: Evaluar el desempeño del test de GM-VClia en muestras de suero y lavado bronquioalveolar (LBA) procesadas previamente con el kit Platelia™ Aspergillus EIA (GM-Plat). Materiales y Métodos: Se estudiaron 56 muestras de suero y 40 de LBA, correspondientes a un total de 59 pacientes (algunos con determinación de galactomamano en ambas muestras) con enfermedades pulmonares, hematológicas, LES, Covid-19 y tumores, entre otros. Trece pacientes tuvieron aspergilosis invasora (1 probada y 12 probables). Resultados: La correlación entre ambos métodos para suero y LBA fue r = 0,8861 p < 0,0001 y r = 0,6368 p < 0,001, respectivamente. Hubo una concordancia global de 67,7% (65/96), siendo de 85,7% (48/56) en sueros y 42,5,0% (14/49) en LBA. Al subir el punto de corte en LBA por GM-VClia la concordancia aumentó a 85,7%. Conclusiones: Se observó una mayor correlación y concordancia en sueros que en LBA. El kit GM-VClia presentó una mayor sensibilidad y valor predictor negativo (VPN), que el kit GM-Plat. Las desventajas de GM-VClia, la constituyen la categoría “dudoso”, que dificulta la interpretación y que, con los puntos de corte actuales en LBA, la correlación con GM-Plat es menor. Las ventajas son su mayor sensibilidad, facilidad de procesamiento y una mayor rapidez en los resultados.