ABSTRACT Objective To assess the cytotoxicity of products subsequent to a cleaning process based on a validated standard operating procedure (SOP), and a final rinse with different types of water: tap, deionized, distilled, treated by reverse osmosis and ultra-purified. Method This was an experimental and laboratory study. The sample consisted of 130 hydrodissection cannulas, 26 per experimental group, characterized according to type of water used in the final rinse. The samples were submitted to internal and external contamination challenge with a solution containing 20% defibrinated sheep blood and 80% of sodium chloride 0.9%. Next, the lumens were filled with a ophthalmic viscosurgical device, remaining exposed for 50 minutes, and then were processed according to the validated SOP. Cytotoxicity was assessed using neutral red uptake assay. Results No cytoxicity was detected in the sample extracts. Conclusion The samples did not display signs of cytotoxicity, regardless of final rinse quality. The results obtained were reached by using only a validated cleaning operating procedure, based on the scientific literature, and on official recommendations and related regulation. SOP, SOP , (SOP) tap deionized distilled ultrapurified. ultrapurified ultra purified. purified ultra-purified study 13 cannulas 2 group 20 80 09 0 9 0.9% Next device 5 minutes assay extracts quality literature regulation (SOP 1 8 0.9 0.
RESUMEN Objetivo Evaluar la citotoxicidad de productos sometidos a contaminación desafío, limpieza siguiendo procedimiento validado y enjuague en diferentes tipos de agua: de caño, desionizada, destilada, tratada por ósmosis inversa y ultrapurificada. Método fueron utilizados 130 canulas de hidrosección, 26 por grupo experimental, caracterizadas por el tipo agua utilizada en el último enjuague. La muestras fueron contaminadas interna y externamente por una solución con 20% sangre de carnero desfibrinado y 80% de Cloruro de Sodio a 0,9%. Luego el lumen fue cubierto por la solución viscoelastica, permaneciendo en contacto con el contaminante 50 minutos y posteriormente procesados, siguiendo orientaciones del procedimiento padrón validado. El ensayo de citotoxicidad se realizó mediante la incorporación del colorante vital rojo neutro. Resultados ausencia de toxicidad en extractos de las muestras. Conclusión no hubo toxicidad en las muestras, independiente del agua utilizada en el último enjuague. Los resultados fueron alcanzados gracias al uso del procedimiento operacional padrón de limpieza validado, embasado en literatura científica, recomendaciones oficiales y en legislación relacionada. desafío caño desionizada destilada ultrapurificada 13 hidrosección 2 experimental 20 80 09 0 9 0,9% viscoelastica 5 procesados neutro científica relacionada 1 8 0,9 0,
RESUMO Objetivo Avaliar a citotoxicidade de produtos submetidos à contaminação desafio, limpeza baseada em procedimento operacional padrão (POP) validado e enxágue final em diferentes tipos de água: de torneira, deionizada, destilada, tratada por osmose reversa e ultrapurificada. Método Estudo experimental e laboratorial. Foram utilizadas como amostras 130 cânulas de hidrodissecção, 26 por grupo experimental, caracterizados, de acordo com a água utilizada no último enxágue. As amostras foram submetidas à contaminação desafio interna e externamente por uma solução contendo 20% sangue de carneiro desfibrinado e 80% de Cloreto de Sódio a 0,9%. Em seguida, tiveram o lúmen preenchido por solução viscoelástica, permanecendo em contato com o contaminante por 50 minutos, sendo então, processadas, de acordo com um POP validado. A citotoxicidade foi avaliada pela captura do corante vital vermelho neutro. Resultados Ausência de citotoxicidade nos extratos das amostras. Conclusão As amostras não demonstraram citotoxicidade, independentemente da qualidade de água utilizada no último enxágue. Os resultados apresentados puderam ser alcançados unicamente por meio do uso de um procedimento operacional padrão de limpeza validado, baseado em literatura científica, em recomendações oficiais e na legislação relacionada. (POP torneira deionizada destilada ultrapurificada laboratorial 13 hidrodissecção 2 caracterizados 20 80 09 0 9 0,9% seguida viscoelástica 5 minutos então processadas neutro científica relacionada 1 8 0,9 0,