OBJECTIVE: To determine the validity of the shake test for detecting freeze damage in aluminium-based, adsorbed, freeze-sensitive vaccines. METHODS: A double-blind crossover design was used to compare the performance of the shake test conducted by trained health-care workers (HCWs) with that of phase contrast microscopy as a "gold standard". A total of 475 vials of 8 different types of World Health Organization prequalified freeze-sensitive vaccines from 10 different manufacturers were used. Vaccines were kept at 5 °C. Selected numbers of vials from each type were then exposed to "25 °C and "2 °C for 24-hour periods. FINDINGS: There was complete concordance between HCWs and phase-contrast microscopy in identifying freeze-damaged vials and non-frozen samples. Non-frozen samples showed a fine-grain structure under phase contrast microscopy, but freeze-damaged samples showed large conglomerates of massed precipitates with amorphous, crystalline, solid and needle-like structures. Particles in the non-frozen samples measured from 1 μm (vaccines against diphtheria-tetanus-pertussis; Haemophilus influenzae type b; hepatitis B; diphtheria-tetanus-pertussis-hepatitis B) to 20 μm (diphtheria and tetanus vaccines, alone or in combination). By contrast, aggregates in the freeze-damaged samples measured up to 700 μm (diphtheria-tetanus-pertussis) and 350 μm on average. CONCLUSION: The shake test had 100% sensitivity, 100% specificity and 100% positive predictive value in this study, which confirms its validity for detecting freeze damage to aluminium-based freeze-sensitive vaccines.
OBJETIVOS: Determinar la validez de la prueba de agitación para la detección de daños por congelación en las vacunas adsorbidas en adyuvantes de aluminio sensibles a la congelación. MÉTODOS: Se empleó un proyecto cruzado de doble ciego para comparar el rendimiento de la prueba de agitación realizada por personal sanitario formado con el de la microscopía de contraste de fases como «criterio de valoración de referencia». Se empleó un total de 475 viales de ocho tipos diferentes de vacunas sensibles a la congelación y precalificadas por la Organización Mundial de la Salud, procedentes de 10 fabricantes distintos. Las vacunas se mantuvieron a 5 º C. Un número determinado de viales de cada tipo se expuso a temperaturas de " 25 º C y " 2 º C durante períodos de 24 horas. RESULTADOS: Se observó una coincidencia total entre el personal sanitario y la microscopía de contraste de fases a la hora de identificar los viales dañados por congelación y las muestras no congeladas. Las muestras no congeladas presentaron una estructura de grano fino al observarlas mediante un microscopio de contraste de fases, mientras que las muestras dañadas por congelación mostraron grandes conglomerados de precipitados concentrados con estructuras amorfas, cristalinas, sólidas y aciculares. Las partículas de las muestras no congeladas medían entre 1 μm (vacunas contra difteria-tétanos-tos ferina, Haemophilus influenzae tipo b; hepatitis B, difteria-tétanos-tos ferina-hepatitis B) y 20 μm (vacunas contra la difteria y el tétanos, aisladas o combinadas). Por el contrario, los agregados de las muestras dañadas por congelación medían hasta 700 μm (difteria-tétanos-tos ferina), con una media de 350 μm. CONCLUSIÓN: El test de agitación de este estudio tuvo una sensibilidad del 100%, una especificidad del 100% y un valor predictivo positivo del 100%, lo que confirma su validez para detectar los daños por congelación de las vacunas con adyuvantes alumínicos sensibles a la congelación.
OBJECTIF: Déterminer la validité de l'épreuve d'agitation pour la détection de dommages occasionnés par la congélation aux vaccins adsorbés aluminiques sensibles à la congélation. MÉTHODES: Un essai croisé en double aveugle a été utilisé pour comparer l'interprétation de l'épreuve d'agitation conduite par des professionnels de la santé formés à cet effet avec celle obtenue par microscopie en contraste de phase en tant qu'«étalon-or». Au total 475 flacons de 8 différents types de vaccins pré-qualifiés par l'OMS, sensibles à la congélation et provenant de 10 fabricants différents, ont été utilisés. Les vaccins ont été conservés à 5 °C. Un nombre choisi de flacons de chaque type a alors été exposé à une température comprise entre -25 °C et -2 °C pour des périodes de 24 heures. RÉSULTATS: Il a été observé une totale concordance des résultats obtenus par les professionnels de la santé et ceux de la microscopie en contraste de phase dans l'identification des flacons endommagés par la congélation et des échantillons non congelés. Les échantillons non congelés présentaient une structure à grains fins en microscopie à contraste de phase, alors que les échantillons endommagés par la congélation présentaient de gros agglomérats de précipités accumulés présentant des structures amorphes, cristallines, solides et en aiguille. En ce qui concerne les échantillons non congelés, les particules mesuraient entre 1 μm (vaccins contre diphtérie-tétanos-coqueluche; Haemophilus influenzae type b; hépatite B; diphtérie-tétanos-coqueluche-hépatite B) et 20 μm (vaccins contre diphtérie et tétanos, simples ou combinés). Les agrégats des échantillons endommagés par congélation mesuraient quant à eux jusqu'à 700 μm (diphtérie-tétanos-coqueluche) et 350 μm en moyenne. CONCLUSION: Dans cette étude, l'épreuve d'agitation du vaccin a présenté une sensibilité de 100 %, une spécificité de 100 % et une valeur prédictive positive de 100 %, ce qui confirme sa validité pour la détection de dommages occasionnés par la congélation aux vaccins adsorbés aluminiques sensibles à la congélation.